RehabTools
2017 · article · Scopus-confirmed

The concurrent validity and intrarater reliability of the Microsoft Kinect to measure thoracic kyphosis

June Quek · Sandra Brauer · Julia Treleaven · Ross A. Clark

International Journal of Rehabilitation Research

Measurement

Abstract

This study aims to investigate the concurrent validity and intrarater reliability of the Microsoft Kinect to measure thoracic kyphosis against the Flexicurve. Thirty-three healthy individuals (age: 31±11.0 years, men: 17, height: 170.2±8.2 cm, weight: 64.2±12.0 kg) participated, with 29 re-examined for intrarater reliability 1-7 days later. Thoracic kyphosis was measured using the Flexicurve and the Microsoft Kinect consecutively in both standing and sitting positions. Both the kyphosis index and angle were calculated. The Microsoft Kinect showed excellent concurrent validity (intraclass correlation coefficient=0.76-0.82) and reliability (intraclass correlation coefficient=0.81-0.98) for measuring thoracic kyphosis (angle and index) in both standing and sitting postures. This study is the first to show that the Microsoft Kinect has excellent validity and intrarater reliability to measure thoracic kyphosis, which is promising for its use in the clinical setting.

Abstract source ↗
Cite this publicationDownload Plain text

Citation fields reflect the indexed metadata. Check author formatting and publication details against the source before submission.

Record provenance

Indexed record: W2621817976. Matched to Scopus author 55279534900.

Scopus record ↗

Metadata retrieved 2026-09-20. Topic labels are navigation aids derived from title and abstract text.