Reliability of maximal isometric knee strength testing with modified hand-held dynamometry in patients awaiting total knee arthroplasty: useful in research and individual patient settings? A reliability study.
This note indexes a cited or audit-added source. It may be an original study, review, guideline, form or methods reference. Treat the specific design and the evidence below as authoritative; a source note is not automatically a primary study.
Bibliographic record from the original reports
Access labels embedded in these original reference strings are historical. Current access and exact checked components are stated separately below.
Koblbauer IF, Lambrecht Y, van der Hulst ML, Neeter C, Engelbert RH, Poolman RW, et al. Reliability of maximal isometric knee strength testing with modified hand-held dynamometry in patients awaiting total knee arthroplasty: useful in research and individual patient settings? A reliability study. BMC musculoskeletal disorders. 2011;12:249. DOI 10.1186/1471-2474-12-249 Source examined: Complete original body including tables.
Koblbauer IF, Lambrecht Y, van der Hulst ML, Neeter C, Engelbert RH, Poolman RW, et al. Reliability of maximal isometric knee strength testing with modified hand-held dynamometry in patients awaiting total knee arthroplasty: useful in research and individual patient settings? A reliability study. BMC musculoskeletal disorders. 2011;12:249. DOI 10.1186/1471-2474-12-249 Source examined: Complete article and tables; preoperative subgroup.
Conditions: COND-KOA Knee osteoarthritis, COND-TKA Total knee arthroplasty
Scoped audit evidence
These are source- and component-level audit summaries as of 3 October 2026. Access and comparison scope can differ by report; none certifies every result or the underlying raw data.
Knee osteoarthritis
- access: fulltext_xml
Knee osteoarthritis — strength
- current access: fulltext_xml
- verification depth: tables
Methods and Table 2 confirm the sample, fixation, force ramp, units and all quoted reliability/error values. The small retest subgroup, variable interval, fixed order and outlier exclusion are correctly retained.
Source locator: Methods and Tables1-2
Limitations: Source-group appraisal. A supported source group does not turn every generic clinical recommendation into a validated decision rule.
Total knee arthroplasty
- access status: full_body_retrieved
32 interrater 13 retest,mean 11 daysrange 2–27;mean 3 trials labeledICC(2,1);N/kg affected extension.96,SEM.21,SDD.58 and 19.0/31.4% match. Preoperative applicability correctly separated.
Source locator: Methods design/protocol/statistics; Results Table 2
Limitations: Body extraction may omit images, equations or supplements. Metadata publication year may represent online-first date, not issue year. Article access does not prove every table/figure/supplement was examined. Use the claim ledger and explicit source locators; do not count this inventory as a complete assertion-level audit.
Reports citing this source
- Strength muscle power and rapid force production assessment in knee osteoarthritis · reference 1 · original reference
- Strength muscle power and rapid force production assessment in knee osteoarthritis · reference 1 · original reference
- Strength power and rapid force assessment after total knee arthroplasty · reference 10 · original reference