Marsan T, Clauzade S, Zhang X, Grandin N, Urman T, Linton E, et al. REEV SENSE IMUs for Spatiotemporal Gait Analysis in Post-Stroke Patients: Validation Against Optical Motion Capture. Sensors (Basel, Switzerland). 2026;26(2). DOI 10.3390/s26020667
This note indexes a cited or audit-added source. It may be an original study, review, guideline, form or methods reference. Treat the specific design and the evidence below as authoritative; a source note is not automatically a primary study.
DOI: 10.3390/s26020667
Bibliographic record from the original reports
Access labels embedded in these original reference strings are historical. Current access and exact checked components are stated separately below.
Marsan T, Clauzade S, Zhang X, Grandin N, Urman T, Linton E, et al. REEV SENSE IMUs for Spatiotemporal Gait Analysis in Post-Stroke Patients: Validation Against Optical Motion Capture. Sensors (Basel, Switzerland). 2026;26(2). DOI 10.3390/s26020667 Source examined: Full text.
Conditions: COND-STROKE Stroke
Scoped audit evidence
These are source- and component-level audit summaries as of 3 October 2026. Access and comparison scope can differ by report; none certifies every result or the underlying raw data.
Stroke gait — gait
- current access statement: Current audit access: full_body_qualification. Prior claims of licensed reading are not re-certified merely from present availability.
- source access: full_body_qualification
Methods and Results confirm 20 enrolled and 14 analysed after excluding six slow walkers, including all walker users. Temporal agreement is stronger than spatial agreement, but differences beyond ±1.96 SD were removed before final accuracy analyses.
Source locator: audit evidence record E039,E045,E054–066; Tables2–4
Limitations: The study trims differences outside ±1.96 SD before final analysis. This can bias retained measurement-error estimates downward. Describing trimming as standard Bland–Altman methodology is misleading.
Linked audit findings
Reports citing this source
- Gait assessment and prognosis after stroke · reference 50 · original reference
Update record
- 2026-10-03: Created from the supplied reports and final scoped audit. New evidence should be appended with source version, access date, exact locator and affected report links.
Updating the wiki · Evidence and status guide
Linked knowledge pages
- TEST-WEARABLEGAIT Wearable and instrumented gait